2. Go to the Network screen

a. Touch the Settings tab.
b. Touch the Device tab.
c. Touch the Network button.

3. Fill in the screen.

If DHCP is enabled for the network, select the DHCP checkbox.

If DHCP is not enabled, fill in the fields: Network IP address, Subnet mask, and

Default gateway.

4. Touch the Test with ping button.
5. Connect to the network.

a. In the Host Address field, enter the host IP address.
b. Touch the Ping Test button. The electrocardiograph pings the host.

6. Choose one of these to specify how to get DCP server IP information:

Obtain server IP information automatically: Select this option if you want the

device to get the IP address, or if you don’t know the server’s IP address.

DCP IP address: Fill in this field if your DCP server IP address is available.

Fill in the UDP broadcast port field.

7. Touch Test connection.
8. Fill in the Remote file server address, User ID, and password.
9. Touch Test remote folder.

The electrocardiograph connects to the remote file server. The electrocardiograph is

ready to send tests to the remote file server. (To set up the workstation, see the

instructions that came with the workstation connectivity software.