Protocol Systems, Inc.
Adjust Limits with STAT SET
When you want to quickly set all alarm limits, the Propaq Encore can calculate
new alarm limits using the patient’s current values. Press the
STAT SET
button in
the Alarms Menu (
SETUP
>
ALARMS
>
STAT SET
). The Propaq Encore activates all
alarms and calculates the limits for all monitored vital signs, except apnea delay.
The monitor performs a mathematical function (addition, subtraction, or
multiplication) on the current value of the vital signs to arrive at the new limits.
The formulas for these calculations are shown in Table 3.
Table 2—Propaq Encore Audible Alarm Indications
Alarm Condition
Tone
Patient alarm for most channels
Continuous ON
SpO
2
alarm
Continuous ON; lower in
pitch
Apnea alarm
ON for 1 second,
OFF for 1 second
Equipment alert
ON for 1 second,
OFF for 4 seconds
Intl Update Doc.fm5 Page 40 Monday, April 29, 1996 3:30 PM