Propaq Encore Version 1.20
43
English
Real-Time ECG Analog/Defib Sync
Impedance Pneumography (RESP)
Invasive Pressure
Pulse Oximetry (SpO
2
)
Table 5—Real-Time ECG Analog / Defib Sync Signals
Characteristic
Specification
Real-Time ECG Output
Range = ± 6V minimum, centered about 0V, Gain = 1000X,
noninverting for Lead II, inverting for all other leads, delay
<3 msec, 0.05 to 100 Hz, going to -5.9V ± 5% during ECG
lead fail. V lead has no Real-Time analog output.
Table 6—Impedance Pneumography RESP Specifications
Characteristic
Specification
Cardiovascular artifact
rejection (CVA)
Presence of CVA is detected automatically. Breaths will be
picked in the presence of CVA unless the Breath Rate is
within 5% of the heart rate or a multiple or sub-multiple of
the heart rate.
Table 7—Invasive Pressure Specifications
Characteristic
Specification
Bandwidth
Digital filtered, DC to 20 Hz
Table 8—SpO
2
Specifications
Characteristic
Specification
Sensor Compatibility
Compatible only with NELLCOR sensors listed in the
Prod-
ucts & Accessories
booklet.
Alarm Hold-off
10 seconds; reset if the sensor reports levels within limits
before 10 seconds elapses.