Spo2 nellcor sensor error codes, Platform spo2 wui software errors and recovery – Welch Allyn Connex Vital Signs Monitor (CVSM) Calibration/Verification Module Notes - Quick Reference Guide User Manual

Page 3

Advertising
background image

SpO2 Nellcor sensor error codes

Nellcor Errors (Section 3.1.1 of Nell-1 Host Controller Interface Specification [6] - see references)

Recovery
number

Description

Recovery

0x01

Irrecoverable malfunction.

Replace Nell-1 board.

0x02

Irrecoverable malfunction.

Replace Nell-1 board.

0x23

Software error; reset in progress.

Replace Nell-1 board if no communication for 15

seconds.

0x45

Communication error

Contact application developer.

0x67

Defective/bad sensor

Fix sensor.

0xDC

Communication error with sensor

Replace sensor.

0xAB

Recoverable error that needed to be

logged.

Not required.

Action taken for these errors are tabulated in the table below:

Platform SpO2 WUI software errors and recovery

Nellcor recovery
number

Masimo board
failure code

Software generated
error

Host controller
recovery action

0x01
0x02

All

BOARD_FAIL

Display a message to

replace the SpO2 board.

0x23

N/A

BOARD_FAIL_STANDBY Wait 15 seconds for the

communication to restart.

If not, display error to

replace the board.

0x67
0xDC

Upon detection of a

malfunctioning probe.

BAD_PROBE

Display a message

indicating that a

malfunctioning probe is

attached.

0x45
0xAB

N/A

None

N/A

N/A

N/A

WRITE_CONFIG

Check CSpO2CData object

used in the

WRITE_CONFIG command

(see explanation below).

Some configuration may be set by sending the WRITE_CONFIG command. However, if the

configuration specified receives an ACK message, but it fails to apply, the software reports a

WRITE_CONFIG error. This is necessary because it may take up to 2 seconds for the configuration

setting to apply and the module should continue to communicate with the host during this time.

Therefore, the WRITE_CONFIG command, if replied with ACK, fails to apply, a WRITE_CONFIG error

is reported.

Advertising