Module verification – Welch Allyn Connex Vital Signs Monitor (CVSM) Calibration/Verification Module Notes - Quick Reference Guide User Manual

Page 7

Advertising
background image

Module verification

1. Click within the module row, and click Verify to activate the test sequence for the module.
2. Enter information about any replacement parts you used during service of the product. After

entering the part information, click Begin to start the tests.

3. Follow the onscreen prompts in the right side pane of the test window to progress through the

test sequence.
The left detail pane shows the order of the required tests and tracks the test sequence

progress.

As the Welch Allyn Service Tool application completes each test, the pass/fail indicator

changes to green for pass or red for fail, and the application proceeds to the next test.

To terminate the active test sequence at any time, click Abort. To terminate all tests,

clickAbort All.

After the Welch Allyn Service Tool application completes all the tests, it displays the pass/fail

test results. After you have finished reviewing the test results, click Close.

Advertising