SpO2 error codes

NOTE:

The Masimo SpO2 module does not provide error codes.

To find the Nellcor error code:

1. Go to the tab of the device you are working on and select the Nellcor SpO2 module.

2. Select the View log button in the user information section of the page. A .txt document opens

with the last Nellcor SpO2 error code.

Post error summary

The following table summarizes the error conditions, actions taken by the module, and actions that

the host should take upon detecting these conditions. The conditions are detected by the module in

the order listed in the table. Only one condition is detected at a time.

Error condition

Host controller action to detect
error condition

Device action

RAM test failure

Start host side WACP application to

communicate with module

Device halts execution, no WACP

over USB communication possible

with module.

ROM test failure

Request module data.

Device does not execute software,

hence sending the incorrect GUID for

request for module data.

ATE test failure

Request POST errors.

POST failure status message with

POST_ATE_FAIL or POST_FACTORY

is reported as a response to request

for POST errors.

SpO2 module

communication failure

Request POST errors.

POST failure status message with

POST_SENSOR or WRITE_CONFIG is

reported as a response to request for

POST errors.

SpO2 Masimo sensor error codes

Table 22 of MS-11 and MS-13 Board Communication Protocol [6] - see references.

Board failure codes

Recover

All

Replace MS-2011 board.