Installation guide

Configuration

21

To configure the plug-in for automatic ECG importing

This procedure configures the CardioPerfect program to import ECG tests into the
CardioPerfect database automatically when new SCP files appear in the specified folder.

1.

Find the TestFileImporter entry in CPSAPI.config.

Example:

<TestFileImporter name=”SCP-ECG Test Importer” enabled=”1”
type=”CPWS.Plugins.TestFileImporter.SCPECGTestFileImporter,CPWS.Plugins” sourcefolder=”specify import folder
here” />

2.

Modify the enabled and sourcefolder attributes as desired.

3.

Restart the CPSAPI service.

a.

Go to Control Panel > Administrative Tools > Services.

b.

Right-click CardioPerfect Services.

c.

Select Restart.

Attribute

Description

Possible Values

enabled

Specifies whether CPSAPI should load and execute the plug-in.

0 (Off), 1 (On)

sourcefolder

Specifies the import folder for incoming SCP-ECG files.
Caution: The destination and source (export and import) folder names
must be unique. If these folder names are identical, the system will
not operate properly.

Any valid Windows path