Brookfield DV-III Ultra Rheometer User Manual

Page 100

Advertising
background image

Brookfield Engineering Laboratories, Inc.

Page 100

Manual No. M98-211-E0912

Universal Spindle Ranges (cont.)

Spindle

Entry Code

SPINDLE RANGE COEFFICIENT

Rheometer Series

LV

RV

HA

HB

Spiral

70

98,400

1,050,000

2,100,000

8,400,000

ULA

00

600

6,400

12,800

51,200

DIN-81

81

3,420

36,500

73,000

292,000

DIN-82

82

3,420

36,500

73,000

292,000

DIN-83

83

11,340

121,300

242,600

970,400

DIN-85

85

1,144

12,200

24,400

97,600

DIN-86

86

3,420

36,500

73,000

292,000

SC4-14

14

117,200

1,250,000

2,500,000

10,000,000

SC4-15

15

46,880

500,000

1,000,000

4,000,000

SC4-16

16

120,000

1,280,000

2,560,000

10,240,000

SC4-18

18

3,000

32,000

64,000

256,000

SC4-21

21

4,688

50,000

100,000

400,000

SC4-25

25

480,000

5,120,000

10,240,000

40,960,000

SC4-27

27

23,400

250,000

500,000

2,000,000

SC4-28

28

46,880

500,000

1,000,000

4,000,000

SC4-29

29

93,750

1,000,000

2,000,000

8,000,000

SC4-31

31

30,000

320,000

640,000

2,560,000

SC4-34

34

60,000

640,000

1,280,000

5,120,000

SC4-37

37

234,000

250,000

500,000

2,000,000

CP/CPE/CPA-40

40

307

3,270

6,540

26,160

CP/CPE/CPA-41

41

1,151

12,280

24,560

98,240

CP/CPE/CPA-42

42

600

6,400

12,800

51,200

CP/CPE/CPA-51

51

4,854

51,780

103,560

414,240

CP/CPE/CPA-52

52

9,300

99,220

198,440

793,760

V-71

71

2,456

26,200

52,400

209,600

V-72

72

10,404

111,000

222,000

888,000

V-73

73

50,146

535,000

1,070,000

4,280,000

V-74

74

508,954

5,430,000

10,860,000

43,440,000

Note:

RV spindle ranges are calculated with the RV Guardleg in use.

LV spindle ranges are calculated with the LV Guardleg in use.

Example 1:

Determine the full scale viscosity range of the LV3 spindle running on an RV series

Rheometer at 45 RPM.

Spindle Speed = 45 RPM

Advertising