Viscometer Will Not Return To Zero

✓

Rheometer is not level

• Adjust the laboratory stand

✓

Pivot point or jewel bearing faulty

• Perform an Oscillation Check*

✓

Manually lift the spindle nut and rotate counter clockwise until the display shows

20% - 30% deflection.

✓

Release the spindle nut.

✓

The % display should return to 0.0%

± 0.1%.

• Perform calibration check
• Contact Brookfield Engineering Laboratories, Inc. or your Brookfield dealer for repair.

* This procedure does not apply to instruments with ball bearing suspension (See Section I.7).

❑

Display Reading Will Not Stabilize

✓

Special characteristic of sample fluid. There is no problem with the rheometer.

• Refer to Appendix C

✓

Check for erratic spindle rotation

• Verify power supply
• Contact Brookfield Engineering Laboratories, Inc. or your Brookfield dealer for repair.

✓

Bent spindle or spindle coupling.

• Contact Brookfield Engineering Laboratories, Inc. or your Brookfield dealer for repair.

✓

Temperature fluctuation in sample fluid

• Use temperature bath control.

❑

No Recorder Response

✓

Be sure the rheometer is not at ZERO reading.

✓

Be sure the recorder is ON and not on STANDBY.

✓

Verify the range settings.

✓

Check cable leads for clean connection.

❑

Recorder Pen Moves In Wrong Direction

✓

Output polarity reversed

• Reverse leads