Brookfield DV-III Ultra Rheometer User Manual

Page 94

Advertising
background image

Brookfield Engineering Laboratories, Inc.

Page 94

Manual No. M/98-211-B0104

Brookfield Engineering Laboratories, Inc.

Page 94

Manual No. M98-211-E0912

Note:

Continuous print mode may be used in conjunction with Timed Stop to

print data throughout the time period. Printing will occur only during

the test.

S

TIMED AVERAGE

The DV-III Ultra will collect a specified number of data points over a period of time and present

the average reading (arithmetic average). Time, speed and the number of data points are entered

using the number keys and

OPTIONS/TAB

key. The test will begin immediately upon the press of

the

ENTER

key. Non-averaged data will be displayed during the test.

The test will end when the specified time interval has elapsed. The averaged data point will be

shown as displayed in Figure V-7. The

SELECT DISP

key can be used to view the average value of

all measurement parameters. The

PRINT

key can be used to send a single averaged dataline to the

printer.

DATA AVERAGING

300 READINGS WERE

AVG'D OVER 02:30 M:S

AVERAGE TORQUE=90.3

Figure V-7

Note:

Printing during a Timed Average test will show non-averaged data.

Averaged data is available only at the conclusion of the test. Printing the

averaged data must be done with the

PRINT

key at the end of the test.