Checkline ETMPX User Manual

Page 13

Advertising
background image

–13 –

To Save The Next Measuring Period:

1. Press the MEM key again. The DISPLAY shows the memory

mode “S” and the next memory number.

6.5 Data Logging in Mode F (Fast)

1. Turn on the tension meter.
2. Memory Mode C set as described in Sec. 6.1
3. Insert process material and begin measuring as described in Sec.4.0.

To Start Data Logging:

1. Press and hold the MEM key until the DISPLAY shows the

memory mode “F” and the current memory number.

2. Release the MEM key. The tension meter starts logging the data at a rate of 100

readings per second.

While the tension data are stored, the MEM indicator blinks on
the DISPLAY and the currently measured value is displayed.

To Stop Data Logging:

1. To stop data logging, press the MEM key once again. The statistical values are

calculated and stored in the following order:

Average value
Last value
Maximum value (MAX)
Minimum value (MIN)
Minimum peak value (MIN PEAK)
Maximum peak value (MAX PEAK)
Tension Value

- 1 -

Tension Value

- 2 -

Tension Value

- N -

up to 4000 tension values in up to

255

measuring

periods.

The MEM indicator is frozen on the DISPLAY and the current
reading is displayed. The tension meter has changed back to
measuring mode.

To Save The Next Measuring Period:

1. Press the MEM key again. The DISPLAY shows the memory

mode “F” and the next memory number.

L 2

cN

F 1

cN

3

Mem cN

F 2

Mem cN

– 24 –

11.0 SPECIFICATIONS

Calibration

According to SCHMIDT factory procedure

Units of Measure

cN / g, user selectable

Accuracy

± 1% FS* ± 1 digit (typical ± 0.5% FS*)

Overrange

10% FS*, without accuracy guarantee

Overload Protection

200% FS*

Measuring Principle:

Strain gauge bridge

Meas. Roller Deflection

0.5 mm, max

Signal Processing

Digital, 16 bit A/D converter

Damping

Adjustable electronical (averaging)

Measuring Frequency

Approx. 5 kHz internal

Display Update Rate:

2x per second

Display

4-digit LCD, height of digit 11 mm

Memory

Average, last value,

maximum, minimum, MAX

PEAK

, MIN

PEAK

Memory Modes:

4 - for up to 4000 readings

Communication frequency

1 to max. 100 readings/sec

Temperature Coefficient

Gain: less than ± 0.01% FS*/°C

Digital Output Signal

USB (Software Tension Inspect > 50 Measuring

values/sec.)

Temperature Range

10 to 45° C

Air Humidity

85% RH, max.

Auto Power Off

Automatical after approx. 3 min. of non-use

Power Supply

LiPo accumulator (60 h continouse use, 3 ½ h

charging time) and AC adapter 100 ... 240 V AC

with country-specific adapters (EU/USA/UK)

Housing Material

Aluminium profile with plastic outer casing (PVC)

Housing Dimensions

197 mm x 58 mm x 47 mm (L x W x H)

Weight (net /gross):

Approx. 340 g / 1250 g

*FS = Full Scale

V-Groove

Line Speed

m/min ... max.

Roller

Material

Standard

2000

Hardcoated

aluminium

V-Groove

Line Speed

m/min ... max.

Roller

Material

Standard

6000

Oxide

ceramic

ETMX Guide Rollers:

ETMPX Guide Pins:

Advertising
This manual is related to the following products: