Checkline ETMPX User Manual

Page 16

Advertising
background image

–16 –

7.3 Recalling Stored Values in Mode L (Limit)

1. Turn on the tension meter and press the RECALL key, You can exit

recall at any time by pressing the POWER key.

Memory Mode LIMIT

Pos:

1 Pos:

2 Pos:

3 Pos:

4

AVG:

22.0 AVG:

12.0 AVG:

12.0 AVG:

12.0

Last:

23.0 Last:

22.1 Last:

22.1 Last:

22.1

MAX:

24.0 MAX:

22.1 MAX:

22.1 MAX:

22.1

MIN:

21.0 MIN:

5.4 MIN:

5.4 MIN:

5.4

PEAK

MAX

:

26.0 PEAK

MAX

:

28.1 PEAK

MAX

:

28.1 PEAK

MAX

:

28.1

PEAK

MIN

:

19.0 PEAK

MIN

:

1.8 PEAK

MIN

:

1.8 PEAK

MIN

:

1.8

10.8

14.2

14.1

15.4

10.0

19.4

11.2

18.3

7.3

22.9

8.9

17.5

6.1

17.3

10.2

7.8

n

10

10

10

– 21 –

2. Release the RECALL key when the value shown on the Display is fairly stable

(the reading might fluctuate greatly).

3. The DISPLAY shows E –10.

o

1st calibration point:

o

1. Hang a weight which

corresponds to 10% of
the tension range from the
measured material, vertically,
as shown.

2. Press the LEVER down

all the way. Thread the
PROCESS MATERIAL
through the MEASURING
and GUIDE ROLLERS.
Slowly release pressure on
the LEVER until the GUIDE ROLLERS return to their original position.

3. Before starting the calibration, move the instrument slowly up and down to

compensate for possible mechanical friction losses and thus ensure repeatability of
the measurements.

4. Press the RECALL key. As long as the RECALL key is

depressed, the DISPLAY shows a decimal value which is
higher than the first decimal value, e.g., 3500. This decimal
value may vary from instrument to instrument.
Write down the value.

5. Release the RECALL key when the value shown on the

DISPLAY is stable (the reading might fluctuate greatly).
The DISPLAY shows

o

2nd calibration point:

o

1. Repeat steps 1 through 4 above using a weight which corresponds to 40% of the

tension range.

2. Press the RECALL key. As long as the RECALL key is

depressed, the DISPLAY shows a decimal value which is
higher than the second decimal value, e.g., 6000. This decimal
value may vary from instrument to instrument. Write down the
value.

3. Release the RECALL key when the value shown on the

DISPLAY is stable (the reading might fluctuate greatly).
The DISPLAY shows E –70.

E –10

3500

6000

E –40

E –70

MEASURING
ROLLERS

PROCESS
MATERIAL

WEIGHT

2. The DISPLAY blinks showing the average value (AVG)

of the first measuring period (POS: 1) and the
symbol.

3. Press the RECALL key. The DISPLAY blinks, showing

the last value (LAST) of the measuring period.

4. Press the RECALL key. The DISPLAY blinks, showing

the maximum value (MAX) or the measuring period
and the symbol.

5. Press the RECALL key. The DISPLAY blinks, showing

the minimum value (MAX) of the measuring period
and the

symbol.

6. Press the RECALL key. The DISPLAY blinks, showing the

maximum peak value (PEAKMAX) of the measuring period
and the

symbol.

7. Press the RECALL key. The DISPLAY blinks, showing the

minimum peak value (PEAKMIN ) of the measuring period,
the PEAK indicator and the

symbol.

8. Press the RECALL key. The measured values no. 1 — n

of the first measuring period can be recalled.

9. Press the RECALL key. The next measuring period (POS: 2) is shown on the

DISPLAY, starting with the average value (AVG).

22

D

cN

23

D

cN

24

D

cN

21

D

cN

26

Peak cN

19

Peak cN

Advertising