6.0 MEMORY MODE

The tension meter features a
data logger with a memory
capacity for up to 4000
readings, with which you
can store different measuring
periods at one or more
machine positions.

The readings are saved 2x per
second, synchronously with
the display update rate, in all memory modes except the “F” mode in which they are
saved 100x per second. All saved readings and statistics can be shown on the display or
transmitted to a PC (e.g. for further processing in Excel). The memory can be allocated
to different measuring periods, depending on the memory mode.

Memory Mode

S

C

L

F

Meas. periods,
max.

255

255

255

255

Readings, max.

-

4000

4000

4000

Max. no. of rea-
dings per position

-

Any

10

Any

Statistics

Yes

Yes

Yes

Yes

Save readings

-

Yes

Yes

Yes

Memory mode “S” STANDARD (default):

The following values of a measuring
period are calculated and saved at a
rate of 2 readings per second:

Average value,

Last value,
Maximum value (MAX),
Minimum value (MIN),
Minimum peak value (MIN PEAK)
Maximum peak value (MAX PEAK)
Individual readings are not saved.

You can save up to 255 measuring
periods.

Memory mode “C” CONTINUOUS:

The following values of a measuring
period are calculated and all readings are
additionally saved at a rate of 2 readings
per second:

Average value,

Last value,

Maximum value (MAX),

Minimum value (MIN),

Minimum peak value (MIN PEAK)

Maximum peak value (MAX PEAK)

You can save up to 4000 readings, split up
into up to 255 measuring periods.

Memory mode “L” LIMIT:

The following values of a measuring
period are calculated and 10 readings are
additionally saved at a rate of 2 readings
per second:

Average value,
Last value,
Maximum value (MAX),
Minimum value (MIN),
Minimum peak value (MIN PEAK)
Maximum peak value (MAX PEAK)

You can save up to 255 measuring
periods with 10 readings each.

Memory mode “F” FAST:

The following values of a measuring
period are calculated and all readings
are additionally saved at a rate of 100
readings per second:

Average value,
Last value,
Maximum value (MAX),
Minimum value (MIN),
Minimum peak value (MIN PEAK)
Maximum peak value (MAX PEAK)

You can save up to 4000 readings, split
up into up to 255 measuring periods.

NOTE: The selected memory mode remains stored in memory even after the instrument
is switched off.

– 28 –

3.0 USING TENSION INSPECT

1. Double click the TENSION INSPECT ICON

to start the program.

01 Button

SETUP

Section 3.1

02 Button

START/STOP

Section 3.2

03 Choice box SCALE

Section 3.3

04 Choice box SAMPLERATE

Select the sampling rate

05 Button

READ

Section 3.5

06 Button

AUTO

Section 3.3

07 Button

RESET

Section 3.3

08 Display

TENSION

Current reading

09 Display

UNIT

Unit of measure of the readings

10 Display

STATISTICS

Display of statistical values of the

measurement

11 Display

DEVICE TYPE

Display the connected unit

12 Display

TIMESHIFTBUFFER Section 3.3

13 Button

SAVE

Section 3.4

14 Display

TIMESHIFT

Section 3.3

Advertising