Hardy WS100 Configuration Software User Manual

Page 42

Advertising
background image

Chapter 6 - Troubleshooting

6-3

give a combined fail result but not for the specific
load cell that failed. You will still have to determine
which Load Cell failed.

2.

Return to Zero Test (RTZ) - This test
returns a pass/fail result on whether
there is a zero load.

•

If you have an HI 215IT Junction
Box, individual load cells can be
tested.

•

If you do not have an HI 215IT
Junction box the result will be for
the combined load cells in your
weighing system.

3.

Digital Volt Meter (DVM) - This test
returns the voltage outputs from the
load cells.

•

If you have an HI 215IT Junction
Box, the voltage output for each
individual load cell is read.

•

If you do not have an HI 215IT
Junction box the voltage output
read will be for the combined
load cells in your weighing sys-
tem.

4.

System Test - Reads the weight for
the load cells in your system.

•

If you have an HI 215IT Junction
Box, the weight for each individ-
ual load cell can be read as well
as the internal reference.

•

If you do not have an HI 215IT
Junction box the weight read will
be combined weight for all the
load cells in your weighing sys-
tem.

Advertising