CHAPTER 3: STAND-ALONE ANALOG CALIBRATION AND SET UP

SETTING UP CURRENT
OUTPUT OPERATING MODE
(0-20mA, 4-20mA, 20-4mA, and
20-0mA)

This procedure sets up the current output
engine such that zero to maximum input
range results in either a 0-20mA, 4-20mA,
20-4mA, or 20-0mA output range.

Setting Mode
1. Press the DOWN or UP buttons until the
PARAM LEDs look like:

Press the ENTER button.
2. Press the UP or DOWN buttons until
the appropriate current mode/range is
achieved as indicated by the VALUE LEDs
as shown below:
0-20mA

4-20mA

20-4mA

20-0mA

Press the ENTER button.

SETTING UP THE INPUT VOLTAGE
RANGE

This procedure sets up the input voltage
range for all 4 channels simultaneously.

Setting Range
1. Press the DOWN or UP buttons until the
PARAM LEDs look like:

Press the ENTER button.
2. Press the UP or DOWN buttons until
the appropriate input voltage range is
achieved as indicated by the VALUE LEDs
as shown below:
±2V

±1V

±500mV

±75mV

Press the ENTER button.

SETTING UP CHANNEL ENABLE

This procedure enables the desired channel
inputs.
1. Press the DOWN or UP buttons until the
PARAM LEDs look like:

Press the ENTER button.
2. Press the UP or DOWN buttons until
the appropriate channels as represented
by the VALUE LEDs are illuminated where
D19 corresponds to Channel 4, D18
corresponds to Channel 3, etc. Press
the ENTER button when the desired input
channel(s) is configured.

The default is

for Channel 1 only (D22).