Kistler-Morse KM STXPlus User Manual

Page 37

Advertising
background image

APPENDIX B: KISTLER-MORSE

®

PROTOCOL SERIAL COMMANDS

Request from Master

Response from STXplus

>aaRXssr

A0000dddssr

Format

>01RX0Br

A000001859r

Example

‘wX’ Command — Write DSP Filter Factor
This command is sent by the master to set the DSP Filter factor of an STXplus. The range of
acceptable values is 1-100(%). For this example, the master sets the DSP Filter factor for an
STXplus at address 01 to 57(%). Any of the 4 leading zeros may be omitted.

Request from Master

Response from STXplus

>aawX0000dddssr

Ar

Format

>01wX579Cr

Ar

Example

‘RY’ Command — Read DSP Filter Step
This command is sent by the master to request the STXplus to send the DSP Filter step. The
range of this value can be between 0 and 2,147,483,647. The default value is 50. For this
example, the master requests the DSP Filter step for an STXplus at address 01 assuming a
format of 2(X.) which returns 1896. and a checksum of 06.

Request from Master

Response from STXplus

>aaRYssr

Affffffffssr

Format

>01RY0Cr

A1896.06r Example

‘wY’ Command — Write DSP Filter Step
This command is sent by the master to set the DSP Filter step of an STXplus. The range of this
value can be between 0 and 2,147,483,647. For this example, the master sets the DSP Filter step
for an STXplus at address 01 assuming a format of 1(X00.) to 5700.

Request from Master

Response from STXplus

>aawYffffffffssr

Ar

Format

>01wY5700.2Br

Ar

Example

‘RZ’ Command — Read DSP Filter Qualify
This command is sent by the master to request the STXplus to send the DSP Filter qualify. The
range of acceptable values is 2-20. The default value is 3. For this example, the master requests
the DSP Filter qualify for an STXplus at address 01 which returns 10 and a checksum of 51.

Request from Master

Response from STXplus

>aaRZssr

A00000ddssr

Format

>01RZ0Dr

A000001051r

Example

B-18