8-6 Correcting the Internal Mass Value

The GX series balance can correct the internal mass value within the range shown below. This
function corrects the internal mass value to conform to an external weight. The corrected mass value
is maintained in non-volatile memory even if the AC adapter is removed.

The internal mass value is corrected as follows:

Model

Correction reference value

Correction range

GX-200
GX-300

200.000 g

GX-400 400.000

g

GX-600
GX-800

GX1000

500.000 g

-0.020 g to +0.020 g

GX-2000
GX-3000
GX-4000
GX-6100

2000.00 g

-0.20 g to +0.20 g

GX-6000
GX-8000

2000.0 g

-2.0 g to +2.0 g

Example: Using the GX-400








Operation

1 Calibrate the balance using the internal mass (one-touch

calibration). Then, place an external weight and confirm
the value to be corrected.

In the example, the value is off by -0.006 gram and the

correction reference value of the GX-400 is 400 grams. So,
the value is to be corrected by 0.012 gram in 400.000 grams.


2 Press the ON:OFF key to turn off the display.

3 While pressing and holding the PRINT key and the

SAMPLE key, press the ON:OFF key. The balance

displays p5 .



With these keys

held down

Press the key

Correct the internal mass

by +0.012 g in 400.000 g.

Calibrate the balance with

the corrected internal mass.

200 g

Use the same weight

200 g