A&D Weighing GX-8000 User Manual

Page 59

Advertising
background image

57

Transmitting all memory data at one time

Confirm that the “Serial interface ( 5if )” parameters are set properly. For details, refer to “10.
FUNCTION TABLE” and “17. CONNECTION TO PERIPHERAL EQUIPMENT”.

1 Press and hold the PRINT key until reCall is displayed.

2 Press the SAMPLE key to display out .

3 Press the PRINT key to display Sout no with “no” blinking.

5 Press the RE-ZERO key to display out go with “go” blinking.

6 Press the PRINT key to transmit all data using the RS-232C interface.

7 The balance displays Clear when all data is transmitted. Press the CAL key to return to the

weighing mode.

Deleting all memory data at one time

1 Press and hold the PRINT key until reCall is displayed.

2 Press the SAMPLE key several times to display Clear .

3 Press the PRINT key to display SClr no with “no” blinking.

4 Press the RE-ZERO key to display SClr go with “go” blinking.

5 Press the PRINT key to delete all data

6 The balance displays end , then reCall when all data is deleted.

7 Press the CAL key to return to the weighing mode.

Advertising
This manual is related to the following products: