After changing a specimen,
always be sure to update the re-
ference measurement for the new
specimen before proceeding.
Put specimen at
haze
port
Perform
reference
measurement
When specimens have only slight
differences in total transmittance,
you need not update the reference
repeatedly.
Statistical evaluations can also be
performed in Continuous Mode.
The values measured here are
written to the same memory area as
in normal measuring mode.
Switch
on
Statistics
or
Results
Use
<operate>
or the foot switch
to save the value
to the statistics
memory.
Stand-by Mode: To save the lamp,
after a certain interval Continuous
Mode is interrupted and Stand-by
Mode appears on the display. The
measurements are continued when a
significant change occurs in the
beam path (for example, when a
new specimen is put in place).
reference
operate
statistic
on/off
result
on/off
operate