Assembly, 1 unpacking, 2 installation – Boekel Scientific 145951-2 Tissue Baths User Manual

Page 4: Operation, 1 controls and indicator lamp (see figure 1)

Advertising
background image

N2400313 Rev. 0 8/09

3

The Tissue Bath has a water temperature range of ambient + 5

°C to approximately 65°C

under normal operating conditions. The unit has an Orienter Block which has a fixed
temperature of 55

°C ± 3°C and a Dryer Block which has a fixed temperature of 40°C ±

3

°C.

3

.

Assembly

3.1

Unpacking
Remove packing materials carefully, and retain for future shipment or storage of the unit.
Inspect for damage. Complete and return the Warranty Registration Card.
Package should contain 1 each of the following:

• Tissue Floatation Bath

• Power Cord

• Glass Dish

• Warranty Card

• Operating Instructions

3.2

Installation

Place the Tissue Flotation Bath on a flat and stable surface, preferably away from drafts.
Plug the power cord into a power supply that matches the voltage listed on the serial
number label on the rear of the unit. Insure that the surface on which the unit is placed
will withstand the radiated heat produced by typical laboratory equipment.

4.

Operation

4.1

Controls and Indicator Lamp (see figure 1)
Please note that the Dryer and Orienter Blocks are only available on models 145951 and
145951-2.

Power Switch:

Located on the rear of the unit.

Heater Indicator:

Illuminates green when the unit is heating.

Alarm Indicator:

Illuminates when unit reaches an over-temp condition.

Lamp Button:

Used to turn the LED illumination on.

Display:

Displays actual bath temperature in normal operating mode,
or high temperature alarm limit when alarm button is
depressed.

Temp. Button:

Used to activate “Set Temperature Mode”.

Alarm Button:

Used to set over-temp alarm limit when depressed.

Set Button:

Used to confirm mode and temperature changes.

Up Arrow Button:

Used to increment the set temperature.

Down Arrow Button:

Used to decrement the set temperature.

Dryer Block Button:

Used to activate the Dryer Block.

Orienter Block Button:

Used to activate the Orienter Block.