Fault diagnosis, Accessories, Technical specifications – Boekel Scientific 145951-2 Tissue Baths User Manual

Page 7

Advertising
background image

N2400313 Rev. 0 8/09

6

5.

Fault Diagnosis

Symptom

Possible Cause

Action Required

1. Unit does not operate

a. Unit not switched on
b. Unit not plugged into

power supply

c. Fuses blown
d. Power supply failure

a. Switch on
b. Plug in, turn on
c. Replace fuses per 8.2
d. Check that other electrical

appliances on the same
circuit are working

2. Bath temperature does not

rise when expected

a. Actual temperature is

higher than set
temperature

b. Temperature control

sensor is not positioned in
the water

c. Temperature control

circuit fault

a. Check set temperature
b. Position sensor in the

water.

c. Have unit checked by

competent person

3. Temperature continues to

rise when not expected

a. Actual bath temperature is

lower than set temperature

b. Temperature control

circuit fault

a. Check set temperature
b. Have unit checked by

competent person

4. Lamp not illuminated

a. Lamp not switched on
b. Lamp failure

a. Press lamp switch
b. Have unit checked by

competent person

6.

Accessories

Part Number Description

904-0013

Glass Dish

7.

Technical Specifications

This equipment is intended for indoor use and will meet its performance figures within the
ambient temperature range of 20

°C to 30°C, with maximum relative humidity of 80% (non-

condensing). Installation Category II (transient voltages). Pollution Degree 2 in accordance
with IEC 664. Suitable for operation at altitudes of up to 6500 feet.

Specifications:

Temperature Set Range:

(Ambient +5

°C) to 65°C

Operating Temperature Range:

20⁰C to 30⁰C

Temperature accuracy:

± 0.3⁰C at 55⁰C

Stability:

± 0.2⁰C

Supply Voltage Range:

145702 and 145951

115V ± 10%, 60 Hz

145702-2 and 145951-2

230V ± 10%, 50/60Hz

Power Rating:

272W

Heating Rate:

Ambient to 55

°C within 45 minutes

Advertising