2 filling the bath, 3 setting the temperature – Boekel Scientific 145951-2 Tissue Baths User Manual

Page 5

Advertising
background image

N2400313 Rev. 0 8/09

4




FIGURE 1

4.2

Filling the Bath
Fill the glass dish with water to a minimum depth of 2cm. Caution: Water should be
placed only in the glass dish. Filling the chamber without the glass dish will damage
the bath and may pose an electrical hazard.
After filling the glass dish, push the
temperature probe (located in the upper right corner of the bath compartment) downward
into the water so that the tip is immersed at least 1cm and is not touching the glass dish.

During operation continue to add water to maintain the 2cm water level.

If it is necessary to remove the glass dish from the unit, first turn the unit off. Make sure
the temperature probe is moved to its full upright position, and then remove the glass
dish.

Caution: If the unit remains on with the temperature probe in its upright position,
or the glass dish is removed or empty, the unit will go into an over-temperature
condition.

4.3

Setting the Temperature


4.3.1 Depress the “Temp.” button on the front panel, the display will change to “tSEt”.

4.3.2 Depress the “Set” button to enter the set point adjust mode.
4.3.3 Use the up/down arrow keys to obtain the desired set temperature. The display

will flash when in the set temperature mode.

4.3.4 When the desired set temperature is achieved, depress the “Set” button. This

value is stored in memory and the unit will retain this value after the unit is
switched off.

4.3.5 As the bath heats up the heater indicator will remain on steady, as the bath begins

to reach equilibrium, the heater indicator will flash.

Heating Indicator

Alarm
Indicator

Display

Up Adjustment
Button

Down
Adjustment
Button

Advertising
This manual is related to the following products: