8. Door Reversal:

Figure 3




8.1 The door on the hybridization oven is reversible for visibility. To change the doors
position simply remove the 3 screws & 3 caps as show in figure 3. Using the same
screws reposition the door to the side desired, be sure that all screws are secured
properly. Place the caps in the unused holes that the hinge previously was mounted to.

9. Maintenance and Service

No routine maintenance is required.

9.1 Calibration

This product has been factory calibrated for optimum performance. A pedigreed
individually serialized thermistor is used to monitor and control temperature in
this unit. These thermistors are +/- 0.1° C absolute accurate, and +/- 0.01°C
reproducible over time (for a given unit). The thermistor R-T table is stored
digitally and is reproducible from unit to unit without the need of an individual
calibration or re-calibration over item. Additionally, thermal modeling of the
instrument allows for the inclusion of digital offsets to the R-T table in FLASH
memory which eliminates the need for individual unit calibration. However, if a
user has a specific need to recalibrate or add a temperature offset, please