contact the “Boekel” customer service department at 1-800-336-6929, to request
a copy of the calibration procedure.

9.2

Cleaning

Disengage power cord prior to cleaning. The floor of the chamber is protected
with a removable stainless steel drip tray. If a spill occurs, use appropriate clean
up procedures as required for radiation or biohazard control. The outer casing
can be cleaned with a cloth dampened with water. Do not submerge or immerse
the Big SHOT in water. The plastic on the front door must not be cleaned with
any abrasive type cleaning agent. This will cause severe scratching of this
material. Before using any cleaning or decontamination method except those
recommended by the manufacturer, users should check with the manufacturer
that the proposed method will not damage the equipment.

9.3

Replacement of Fuses

Disconnect the unit from the power supply.

Remove the plug from the socket in the back of the unit.

Pull back on the fuse drawer catch.

Pull out the fuse drawer (see Figure 4).

Check and replace with the correct fuses if necessary.

The fuses should be 5mm x 20mm:

230402 (115V)

4.0A/250v 3AG SLO-BLO

230402-2 (230V) 4.0A/250v 3AG SLO-BLO

Push the fuse drawer back in. Reconnect unit to the power supply.

Figure 4

OPEN FUSE
HOLDER AT
THIS POINT
BY PULLING
DOWN.

FUSE DRAWER
PULL OUT TO
REMOVE.

POWER

SWITCH

BE SURE THAT THE

PROPER VOLTAGE

APPEARS AT THE

ENTRY WINDOW

WHEN REINSTALLING.