Accessories – Boekel Scientific 230502-2 Little S.H.O.T. III User Manual
Page 8

8
disengage the front mounting point. To install Carousel, reverse procedure.
Warning! Sharp rotating parts. Do not attempt to remove the carousel until
motor has stopped. Remove power from the equipment. Use caution when
removing the carousel since it may be very hot.
4.9
Drip Tray
To access the drip tray for cleaning, it is necessary to remove the carousel (see
section 4.8). Use caution when removing the drip tray since it may be very hot.
4.10 Reducing the risk of a Hybridization Bottle leak
4.10.1 Glass Hybridization Bottles
Avoid over-tightening the cap, which can twist or cause a gap in the O Ring seal.
Turn the cap a few degrees of arc past the point where the cap is seated to the
bottle.
Replace the O Rings before they become worn, flat or cracked. We suggest
replacement after 10 hybridization runs.
4.10.2 50ml Conical Centrifuge Tubes
Be sure to use tubes that can handle the elevated temperatures used in
hybridization. We recommend that you contact your supplier or tube
manufacturer. In general, scientists have found that tubes made from
polypropylene with polyethylene caps provide reliable performance. Tubes made
from polystyrene should be avoided.
When inserting the centrifuge tube into the clamping mechanism, place the
clamp toward the middle of the tube. Pinching the tube near the cap may cause
a leak.
5. Accessories
Model Number
Description
Glass Hybridization Bottles, 2/package
502-0100
35mm X 100mm
502-0150
35mm X 150mm
502-0225
35mm X 225mm
502-0300
35mm X 300mm
Caps and “O” Rings for Glass Hybridization Bottles
505-0001
Blue caps with silicone “O” rings, package of 2
502-0500
Replacement silicone “O” rings for blue caps, package of 10
509-0009
Red caps, package of 2
Nylon Mesh
230415
15cm X 10cm, package of 5
230423
23cm X 23cm, package of 5
Carousels and Clips
230710
Microplate Carousel (holds 4 standard microplates)