3.1.3 Chip Size LED Error Codes

If self test fails, on rare occasions, the chip size LEDs may light to indicate an error. In the list
below, the 8 LEDs are shown with a 0 for unlighted and a 1 for lighted.

00000001: KEYBOARD CIO PortA. CPU cannot communicate.
00000010: KEYBOARD CIO PortB
00000011: TIME QO PortA
00000100: TIME CIO PortB
00000101: FREQ CIO PortA
00000110: FREQ CIO PortB
00000111: Display not ready
00001000: ROM checksum error
00001001: SRAM error
00001010: Stack underflow
00001011: Unimplemented interrupt occurred

00001100: Invalid intermpt vector
00001101: Not used
00001110: Incorrect peripheral serviced
00001111: UART transmit error
00010000: UART other errors ( special receive conditions)
00010001: UART parity error
00010010: DRAM error
00010011 : Overlay error
00010100: Not used
00010101: MAP CIO PortA
00010110: MAP CIO PortB
00010111 : Display memory error