Tests 42 43, Pullo_act_tston, pull1_act_tston, Tests 44, 45 – Fluke 900 User Manual

Page 34: Pullo_act_actstr, pull1_act_actstr, Tests 46,47, Interface buffer shift register subtest, Interface buffer shift register subtest - 1 0

Attention! The text in this document has been recognized automatically. To view the original document, you can use the "Original mode".

Advertising
background image

Seiftest

FLUKE 900 SERVICE MANUAL

The bits in the tesults indicate nonnal pin numbers with the exception of the following:

BYTE 0 - B7 - NAJ

B6- N/U

B5 - Could not clear PAL latches.
B4 - Could not activate aU lines with a single edge.

TESTS 42 43

PULLO_ACT_TSTON, PULL1_ACT_TSTON

These tests walk a rising edge(l) or a falling edge(0) horn the puU-iq>s latches through the
activity circuitiy. The activity circuit is controlled by the TSTON signal.

TESTS 44, 45

PULLO_ACT_ACTSTR, PULL1_ACT_ACTSTR

These tests walk a rising edge(l) or a falling edge(0) fiom the pull-ups latches through the

activity circuitry. The activity circuit is controlled by die ACTSTR signal.

TESTS 46,47

freq

_

bias

_

short

,

freq

_

bias

_

long

These two tests are almost identical, the only difference being that the long test verifies aU
possible values of threshold and the short test only tests at 1 volt increments. The tests are a
collection of several subtests involving frequency and threshold hardware on the High Speed
Board and Input Buffer and they are detiuled below. As they are being done, the left part of
the screen shows timing values from the frequency test and the right side of the screen shows
the highest and lowest differences between expected and measured voltage values.

Many of the subtests require that a voltage on the Interface Buffer be measured. One of

several possible voltages is selected using the analog mux on the input buffer. This feeds the
into the VCO, the output of which is sent to the High Speed Board where the frequency is

measured. The same thing is also done for 0 volts and 2.5 volts to obtain a calibration ratio
that is used to determine the exact voltage of the signal selected.

Interface Buffer shift register subtest

This tests to see if data can be shifted through the shift registers on the Interface Buffer
correctly. There are two banks of registers on the board. Only the puUup registers are
explicitly tested because there is no feedback line from the threshold registers. The

threshold registers are implicitly tested by the other tests. The test procedure is as follows:

a) shift data - msbit = 0

b) read the voltage of the msbit: must be <2V
c) shift data - msbit = 1
d) read the voltage of the msbit: must be >2V

3 - 1 0

Advertising