V. calibration procedures – Galaxy Audio CHECK MATE CM-130 User Manual
Page 8
Advertising

SOUND LEVEL METER
6
V. Calibration Procedures
Using a standard Acoustic Calibrator
(94dB, 1KHz sine wave)
70 dB
90 dB
130 dB
110 dB
100
60
80
40
LEVEL
SLOW
FAST
A / C
MAX
(1) Make the following switch settings.
Display: dBA
Time Weighting: FAST
Measurement Mode: MAX Mode function disable.
Level Range: 80 to 110dB
(2) Insert the microphone housing carefully into the insertion
hole of the calibrator.
(3) Turn on the calibrator and adjust the CAL potentiometer
of the meter until the level display indicates the desired
level. The meter is calibrated before shipment. We
recommend recalibration annually.
Screwdriver
Acoustic Calibrator