Grant Instruments LTC4 User Manual

Page 25

Advertising
background image

LTC4

31115 V1

Operating Manual

Page 24

www.grantinstruments.com











5. Temperature is highlighted, press S to select.

6. Rotate the dial to define the temperature value,

press S to set.


7. Rotate the dial to highlight latching, press S to

select.


8. Rotating the dial will display on/off, press S to

select.

When latching is on, a temperature alarm
continues unless acknowledged by the user even
if the temperature comes back in range

9. Rotate the dial to highlight relay, press S to select.

10. Rotating the dial will display on/off, press S to

select.


11. Rotate the dial to select buzzer, press S to select.

12. Rotating the dial will display on/off, press S to

select.


13. Rotate the dial to save, press S. The display

returns to the alarm settings screen. The
procedure can be repeated to configure a
deviation alarm and for the low alarm function.


14. Rotate the dial to highlight holdoff, press S to

select.


15. Rotate the dial to select a holdoff time (over a

range of 5 to 600 seconds), press S to select.

Rotate the dial to save, press S to save all
settings.

The display will return to the settings menu



16. Press F to return to the home screen




.



°c

High Alarm mode

Alarm Settings

Low Alarm mode

Save

Fixed Temp

Disabled

Holdoff (s)





.



°c

Temperature

Latching

High Alarm

Off

°c



.

On

Relay

Buzzer

On

Save