11.0 Troubleshooting

Symptom

Possible cause

Action required

Unit fails to cool Compressor

overheated causing
internal over-
temperature
thermostat to operate

Wait for compressor to cool, when thermostat
will reset and compressor will switch on again.

Unit failing to
cool below 5°C

Freezing protection
switch is on and is
preventing further
cooling

Switch off freezing protection (see section
5.2.7).

Unit not cooling
at higher
temperatures

Over temperature
protection switch has
tripped

Refrigeration unit may have temperatures in
excess of 100°C. Leave the unit to cool and
reset (see section 5.2.7)

Cooling
performance
reduced

Condenser fins
clogged

Carry out routine cleaning (see section 9.2)

Temp continues
to rise when not
expected

Set temp is higher
than liquid temp

Check that the bath set temperature is correct
(see section 5.2.10).

Set temperature
too restricted

Liquid type set does
not allow required set
point

Change to different liquid type (see section
5.3.2).

Temperature
does not rise
when expected

Set temp is lower
than liquid temp

Check that the bath set temperature is correct
(see section 5.2.10).

Display shows
“High
Temperature
Alarm”

High temperature
warning alarm has
tripped

Check that the bath set temperature is correct
(see section 5.2.10).

Check that high temperature alarm is correct
(section 5.3.4).

Check that the liquid level in the bath is
adequate (see section 5.1.1 for minimum fill
levels).