Completing a calibration – Grant Instruments LTC4 User Manual

Page 28

Advertising
background image

31115 V1

LTC4

Page 27

Operating Manual

www.grantinstruments.com

3. Rotate the dial to toggle between on/off, press S

to save the selection.


The display returns to the settings screen.

4. Press F to return to the home screen.



5.4

Completing a calibration


The TX150 allows up to five temperature points to be calibrated. The calibration menu
can be accessed by simultaneously pressing the F and S buttons for 3 seconds

The calibration temperatures are constrained by the temperature limits of the liquid type
setting. Calibration should be carried out using a traceable reference thermometer with
an accuracy of at least 0.1°C. This thermometer should be held securely in the centre of
the bath or vessel.

Two factory defined calibration points (20˚C and 70˚C) already exist and are displayed in
the calibrate probe menu, these may be recalibrated if required and up to a further three
calibration points added. Calibration points should be chosen to be at critical
experimental temperatures where accuracy is important or at the extremes of the working
range of used temperatures.

The TX150 calculates the temperature at any point using calculated values that pass
through each calibration point. This ensures precision at all critical experimental
temperatures throughout the range in use. In the example below, 4 calibration points
have been used.

User recorded
Temperature

Measured
Temperature

t3

t4

t2

t1

T4

T3

T2

T1



.



°c

Alarms

Settings

Buzzer

Language

Relay

On