Symptom

Possible cause

Action required

Display shows
“Low liquid
Alarm

Liquid level has
dropped below
minimum level

Check that the liquid level in the bath is
adequate

Display shows
“Internal
temperature
probe Fault”

Faulty temperature
probe

Have a competent person check the probe for
an open or short circuit fault or contact Grant.

Display shows
“Internal/External
Temp Diff Too
Big”

There is a
temperature
difference of more
than 5 degrees
between the internal
and external probe

Check the external probe is properly connected
at the rear panel.

Check the external probe is correctly positioned
in the bath liquid and circulation is not
restricted.

If the external probe is not required, select the
internal probe using the Settings menu

Display shows
“Overheating-
Power reduced ”

Heating water at or
near to boiling
without lid

Heating very large
volumes of liquid with
large thermal losses

Add a lid to reduce thermal losses and leave
unit running with the “Overheating” warning
present. If the alarm has not cleared within 1
hour contact Grant.

Add measures to reduce thermal losses and
leave unit r

unning with the “Overheating”

warning present. If the alarm has not cleared
within 1 hour contact Grant.

Display shows
“Service required
01

Faulty fuse(s), relay
or heater element

Have a competent person check the product or
contact Grant.

Display shows
“Service required
02

Pump or propellor is
obstructed
Faulty motor

Remove obstruction.
Have a competent person check the product or
contact Grant

Display shows
“Cross-check
Failure”

PCB fault

Have a competent person check the product or
contact Grant.




For any other errors or service requests, please contact Grant Instruments service
department.