Grant Instruments TXF200 R User Manual

Page 13

Advertising
background image

Equipment:

This type of complex thermal ramping requires the use of either a GR150 or a
GP200 Optima thermostat. Any of the Refrigeration units can be controlled
with the exception of R1/R1L. As this example will reduce the temperature
down to -20°C, we can use any of these units. You will also require the
Communications lead as supplied with the Refrigeration unit.

Liquid:

From the section on liquids you will see that the suitable type is 50% water
and 50% glycol.

4.2.1

Procedure:

Page13

Low temperature bath/circulator
Operating instructions

Version 4 - Jan 2010

Connect the socket end of the Communications lead to the Relay 1 connection on the


rear of the GP200 and the plug end to the communications lead connection ( ) on the
rear of the refrigeration unit.

Fill with liquid and turn on.
‚Press

button on front of thermostat and rotate knob until [

] is

displayed. Press

and rotate knob to display [

] and press

F

S

S.

Selectliquid

LowTempOil

ƒPress

and rotate knob until [

] is displayed, press

. [

] icon is displayed;

rotate the knob until [

] is displayed and press select

F

S

Programs

Program Editor

[

] Rotate knob so that [

] is displayed.

Editprogram 1

1Temp

Phase 1
[

] is displayed. Now press the

button so the temperature is flashing, rotate

the knob to the required temperature of -20°C and press

to select.

1Temp

S

S

Rotate the knob so that [

] is displayed. Press

to adjust, the time will now

flash, rotate the knob to the required time (in this case 40 minutes) and press

to

select.

‚

1Tim

e

S

S

Phase 2
Rotate the knob so that [

] is displayed, press

and [ ] will be

displayed, introducing Phase 2.

1.InsertAfter

S

2

‚Rotate the knob so that [

] is displayed and press

the display will flash, adjust

to -20°C then press

to accept.

2Temp

S,

S

Rotate so that [

] is displayed and press the

button, adjust to 1 hour.

ƒ

2Tim

e

S

Phase 3
Rotate the knob so that [

] is displayed, press

and [ ] will be

displayed, introducing Phase 3.

2.InsertAfter

S

3

‚Rotate the knob so that [

] is displayed and press

to adjust. Adjust to

40°C then press

to accept.

3Temp

S

S

ƒRotate so that [

] is displayed and press the

button to adjust to 40

minutes.

3Tim

e

S

„As this example is heating to above ambient we do not now require the refrigeration
unit to be on. Rotate the knob so that [

] is displayed, by pressing the

button

you can activate the relays. A [ ] next to the numbers [ ] or [ ] indicates whether that
particular relay will be activated throughout Phase 3. Relay 1 needs to be activated,
so press the

button until just the

[ ] is displayed.

3.Relays

*

1

2

*

S

S

1

4