It is the user's responsibility to carry out appropriate decontamination if hazardous material is spilt on or
inside the equipment.

Do not connect to a power supply or switch on before filling the tank.

Take care when topping up or draining, as the liquid in the tank may be very hot or cold.

If the alarm lamp is illuminated do not touch the liquid or the tank base, they may be very hot. Refill
carefully, a hot heater can cause a spattering of very hot water droplets and scalding steam.

Always use a lid or polypropylene spheres when operating above 60°C. Take care when raising and
removing the lid, it may be hot. Steam and hot vapours can cause scalding.

Drain before moving the bath. Before draining allow the liquid to cool to below 50°C.

Use only liquids specified in the operating instructions, within the specified temperature range. If the alarm
lamp is illuminated the liquid temperature may be above its recommended maximum: do not inhale the
vapours given off as they may be toxic; liquids should be safely discarded and replaced.

When pumping to external apparatus, always use connecting pipe or hose suitable for the operating
temperature and liquid used, and check that the pipe connections are secure.

Do not use Silicone tubing when pumping Silicone oil.

Do not disconnect any pipes or hoses while they contain very hot or cold liquid.

Refrigerated circulators are heavy, take care when lifting or moving; empty weights vary from 24kg to 88kg.

Do not touch condenser fins, they are sharp and may cut.

Page 5

Low temperature bath/circulator
Operating instructions

Version 4 - Jan 2010