2. Getting started

2.1

Unit overview

2.2

Unpacking

Optima low temperature circulators provide a source of cooling for many sensitive
analytical procedures. Applications requiring the precision temperature control of
samples include spectrophotometry, viscometry, refractrometry and electrophoresis.
Grant’s range of more powerful chillers are ideally suited for applications requiring the
removal of the mechanical or electrical heat produced in apparatus or machinery,
combined with a higher pressure flow for faster heat removal within restricted systems.

Remove packing materials carefully, and retain for future shipment or storage of the unit.
Packs should contain:

2.3

Installation

To avoid condensation on internal components, allow the unit to reach room temperature
before switching it on.
Let the unit stand in its intended working position for 6 hours before switching on. This is
to allow the oil to drain to the bottom of the compressor.
Position the equipment so that there is free air flow through it. Allow at least 100mm
clearance from obstructions on all sides.

COOLER UNIT

Lid
Mains cable
Communications cable
Interconnecting power cable
Operating instructions
Operating voltage

R1

R1

R2

R3

R4

R5

R1L R2L R3L R4L

240V 240V 240V 240V

240V

115V 115V

115V

115V

Page 6

Low temperature bath/circulator

Operating instructions

Version 4 - Jan 2010