5. Fault diagnosis

Unit does not operate

Temperature does not rise

when expected

Temperature continues to
rise when not expected

Unit fails to cool

Unit not switched on

Unit not plugged into
power supply

Fuse blown in unit or
in plug (UK units only)

Electrical power supply
failure

Overtemperature cut-out
has operated

Set temperature is lower
than liquid temperature

Liquid level cut out has
operated.

Temperature control
circuit fault

Set temperature is higher
than liquid temperature

Compressor over heated
causing compressor internal
overtemperature thermostat
to operate

Temperature control
circuit fault

Switch on

Plug in, switch on

Check and replace

Check that other electrical
appliances on same circuit
are working.

Reset cut-out as described in

Check set temperature

Check liquid level.

Have unit checked by
competent person

Check set temperature

Wait for compressor to cool, when
thermostat will reset and
compressor will switch on again

Have unit checked by
competent person

thermostat Operating Instructions

Symptom

Possible cause

Action required