Fuse ratings
IEC Socket fuse ratings
The XUBA1 and XUBA3 baths are fitted with an IEC socket. This IEC socket is fitted
with two 20mm fuses.
Should these fuses need to be replaced, the replacement fuses should ALWAYS be
anti-surge. This type of fuse is marked with a “T”; e.g. T1.6A.
Fitting fuses of the incorrect rating in the IEC socket will invalidate the warranty. If you
have any doubt, please contact the Grant Service Department
UK Mains plug fuse ratings
As well as the fuses in the IEC socket, the XUBA1 and XUBA3 are fitted with a 5A
fuse in the mains plug.
Should these fuses need to be replaced, the replacement fuses should ALWAYS be
5A.
Fitting fuses of the incorrect rating in the plug will invalidate the warranty. If you have
any doubt, please contact the Grant Service Department
Model number
Fuse rating
XUBA1
1.6A
XUBA3
1.6A
Model number
Fuse rating
XUBA1
5A
XUBA3
5A