Fuse ratings

IEC Socket fuse ratings

The XUBA1 and XUBA3 baths are fitted with an IEC socket. This IEC socket is fitted
with two 20mm fuses.

Should these fuses need to be replaced, the replacement fuses should ALWAYS be
anti-surge. This type of fuse is marked with a “T”; e.g. T1.6A.

Fitting fuses of the incorrect rating in the IEC socket will invalidate the warranty. If you
have any doubt, please contact the Grant Service Department

UK Mains plug fuse ratings

As well as the fuses in the IEC socket, the XUBA1 and XUBA3 are fitted with a 5A
fuse in the mains plug.

Should these fuses need to be replaced, the replacement fuses should ALWAYS be
5A.

Fitting fuses of the incorrect rating in the plug will invalidate the warranty. If you have
any doubt, please contact the Grant Service Department

Model number

Fuse rating

XUBA1

1.6A

XUBA3

1.6A

Model number

Fuse rating

XUBA1

5A

XUBA3

5A