XUBA Series

7

29667 V1

Operating Manual

EN

XUBA3 Control panel instructions

To switch on the ultrasonics:

Turn the TIME dial to the desired time, and the press the SONICS button.

The SONICS button and the SONICS light will glow. Ultrasonic activity will then
commence in the liquid inside the tank and the Timer dial will be heard ticking.

At the end of the timed period, the TIME will click off, the SONICS light will go out, and
the ultrasonic activity in the liquid will stop.

To stop the ultrasonic activity at any time, press the SONICS button while in opera-
tion.

To operate the heater:
Turn the HEATER dial to the desired temperature, and press the HEAT button.
The HEAT button and the HEAT light will glow, and the liquid will start to heat up.
When the set temperature is reached, the HEAT light will go out.
To stop the heater at any time, press the HEAT button while in operation (ie. when the
HEAT light is already on).

NOTE: Ultrasonic activity itself will heat the liquid. This means that the liquid tempera-
ture may rise above the level indicated on the temperature dial.
The temperature dial only controls the heater cut-out temperature.
If in doubt, check the actual temperature with a thermometer.