Environment of use, Installation, Others – Canon CX-1 User Manual

Page 15

Advertising
background image

2. Notes for Using the Instrument

-3-

• Never use disinfecting ethanol, glutaraldehyde or other solvents to clean the cover of the instrument,

except the forehead rest and the chin rest. This could damage the cover of the instrument.

• If the chin rest paper will not be used, be sure to disinfect the chin rest in the same way as the forehead

rest whenever the patient changes.

Environment of use

• Use the camera in an environment that satisfies the following requirements.

Use

Temperature:

10°C to 35°C

Humidity:

30% to 80% RH (no condensation)

Atmospheric pressure:

800 hPa to 1060 hPa

Storage and Transportation

Temperature:

–30°C to 60°C

Humidity:

10% to 60% RH (no condensation)

Atmospheric pressure:

700 hPa to 1060 hPa

• Do not install or store the camera or leave it standing in an adverse environment such as where the

temperature and humidity levels are high. Doing so may cause trouble and/or malfunction. Be sure to
observe the points of Installation and Environment of Use (see page (3)) when selecting the
installation location.

• Dust in the air will not only attach to the objective lens, but will also attach onto the optical parts inside

the instrument. You cannot take a good image when dust is on them. Please keep the room clean.

Installation

• Please ask a Canon representative or distributor to perform installation of this product.
• Please handle this product carefully. The adjustment can be altered if it is subjected to a strong shock or

jolt.

• If this product will be transported in an automobile or shipped a long distance, protective measures need

to be taken for vibrations and shocks. Ask a Canon representative or distributor for more information.

Others

For U.S.A.
Rx only-Caution:
Federal law restricts this device to sale by or on the order of a licensed practitioner.
Intended Use:
The device is intended to be used for taking digital images of retina of human eye.

For European Union
Intended Use:
This medical device is intended to observe image and record retinal fundus through the pupil without
contact with subject’s eye for the purpose of diagnosis by way of producing fundus image information.

Advertising