 Press the TARE button, the acceleration

value due to gravity is displayed.

 Press the TARE key anew.

„UnLoAd“, followed by „LoAd“ is displayed

 Put the adjustment weight (see chap. 1)

carefully in the center of the weighing
plate, „PASS“ will be displayed.

 While the balance carries out a self test,

remove the adjustment weight.

After successful adjustment the balance
automatically returns to weighing mode.

In case of an adjustment error or incorrect
adjusting weight the display will show an
error message; repeat adjustment process.