KERN & SOHN CPB 30K5DM User Manual

Page 17

Advertising
background image

 TARE button, „UnLoAD“ appears
 Press the TARE key anew

„LoAd“ as well as the currently set
adjustment weight appear flashing

either

• Confirm with TARE

or

• In order to change, enter the desired

weight value of the adjustment weight
by the numeric keyboard and confirm
on the TARE button.

In order to achieve high-quality weighing
results in the sense of the measuring
technology, it is recommended to select the
nominal value as high as possible.
We recommend 80 % max.

+

(Example)

 Carefully place adjusting weight in the

centre of the weighing plate

 Wait for stability display, then press the

TARE button

 While the balance carries out a self test,

remove the adjustment weight.

After successful adjustment the balance
automatically returns to weighing mode.

In case of an adjustment error or incorrect
adjusting weight the display will show an
error message; repeat adjustment process.