15 instant help – KERN & SOHN CPB 30K5DM User Manual

Page 48

Advertising
background image

15 Instant help

In case of an error in the program process, briefly turn off the balance and
disconnect from power supply. The weighing process must then be restarted from
the beginning.

Fault

Possible cause

The displayed weight does not glow.

• The balance is not switched on.

• The mains supply connection has been

interrupted (mains cable not plugged
in/faulty).

• Power supply interrupted.

• Batteries are inserted incorrectly or empty

• No batteries inserted.

The displayed weight is permanently
changing

• Draught/air movement

• Table/floor vibrations

• Weighing plate has contact with other

objects.

• Electromagnetic fields / static charging

(choose different location/switch off
interfering device if possible)

The weighing result is obviously
incorrect

• The display of the balance is not at zero

• Adjustment is no longer correct.

• Great fluctuations in temperature.
• Electromagnetic fields / static charging

(choose different location/switch off
interfering device if possible)

CPB-N / CPB-DM-BA-e-1424