6.7.2 Models CPB-DM (verified models)

The adjustment is locked for verified balances. Carrying out adjustment
requires that the seal is destroyed and the adjusting switch is pressed when
turning on the scale. For position of adjusting switch, see chap. 6.9.1.

Attention:
After destruction of the seal the balance must be re-verified by an
authorised agency and a new verification wire/seal mark fitted before it can
be reused for applications subject to verification.


Operation

Indication

 Switch-on balance and during the selftest

actuate the adjustment switch and press
the TARE button

 Use the number keys to enter password:

either

• Default password "0000"

or

• Personal password, enter under

function

[F6 PKn ]

see chap.12

 Confirm with TARE button


CPB-N / CPB-DM-BA-e-1424