6.8 Linearization (non-verified models only)

Linearity shows the greatest deviation of a weight display on the scale to the value of
the respective test weight according to plus and minus over the entire weighing
range.
If linearity deviation is discovered during a testing instrument control, you can
improve this by means of linearization.

• Carrying out linearization is restricted to specialist staff possessing well

acquainted with the workings of weighing scales.

• The test weights to be used must be adapted to the weighing scale’s

specifications; see chapter 3.4 “testing instruments control”.

• Observe stable environmental conditions. Stabilisation requires a

certain warm-up time.

• After successful linearization you will have to carry out calibration; see

chapter 3.4 “Testing instruments control”

Tab. 1: Adjustment points

Adjustment
weight

CPB 6K0.1N

CPB 15K0.2N

CPB 30K0.5N

1.

1 kg

2.5 kg

5 kg

2.

2 kg

5 kg

10 kg

3.

4 kg

10 kg

15 kg

4.

6 kg

15 kg

30 kg

Operation

Display

How to carry out linearization:

 Switch on balance and during self-check

press the TARE key

 Use the numeric keys to enter password

„9999“

 Confirm with TARE button


CPB-N / CPB-DM-BA-e-1424