Instruction manual fh, Sauter gmbh – KERN & SOHN FH 100K User Manual

Page 4

Advertising
background image

Sauter GmbH

Ziegelei 1
D-72336 Balingen
E-Mail: [email protected]

Tel: +49-[0]7433- 9933-0
Fax: +49-[0]7433-9933-149
Internet: www. kern-sohn.com

Instruction Manual

FH

FH_SO-BA-e-1217

4

 To browse through the stored values, please use the

▲ or ▼ keys. (The values will be shown in the upper
display segment)


 By pressing the MEMORY key, the average value of

the stored peak values can be shown (in the upper
display segment)


 To delete every stored value, press the

▼-key in the

AVERAGE-Mode


6.

Configuration of RS 232

SUB-D 9pm

Pin

Signal

Illustration

2

TxD

Output signal

3

RxD

Input signal

5

GND

Ground

6

+1.6 to + 2 V

Over upper limit

7

+1.6 to + 2 V

Lower than lower limit

8

+1.6 to + 2 V

OK

6.1 Output Protocol


RS-232 Parameter
Baud rate: 9600
Data-Bit: 8
Parity: none
Stop-Bit: 1

The measured value is requested by the PC by the ASCII
Sign “9”.

The measured value that comes from the instrument has
this format:

e.g. 0011.70 means -11,70 Newton, if
Newton is the selected unit
||-----|
| |____> the other 6 places describe the
measured value as ASCII-Signs
|_______> the first place describes the
direction of the force (0 = minus = Pressure;
1 = plus = Tension)

or: 1021.15 means +21,15 N (Tension)



7. Warning


Intended use

The instrument you have acquired serves to determine the
measuring value of the material to be measured. It is
intended to be used as a “non-automatic“ instrument, i.e.
the material to be measured is manually and carefully
attached at the instrument. The measuring value can be
read off after a stable measuring value has been obtained.

Inappropriate use

Do not use the instrument for medical measurements. In
the case that small quantities are removed or added to the
material to be measured, incorrect measuring results can
be displayed due to the “stability compensation“ inside of
the instrument. (Example: Slow draining off of liquid from a
container suspended from the instrument). Do not attach a
continuous load. This can damage the measuring unit as
well as the parts, relevant to safety.
Prevent jolts, torsion and oscillation (e.g. by appending
slopingly) of all kinds. Be sure to prevent overloading the
instrument in excess of the stated maximum load (max.),
minus any tare weight that may possibly exist. This could
damage the instrument (risk of breakage).
Important:
• Always make sure that there are no people or materials
below the load that could be injured or damaged!
• The instrument is not suitable for measuring people. Do
not use as baby scales!
• The instrument does not comply with the medical product
law (MPG).
Never operate the instrument in hazardous locations. The
series design is not explosion-proof. Structural alterations
may not be practised to the instrument. This can lead to
incorrect measuring results, faults concerning safety
regulations as well as to destruction of the instrument. The
instrument may only be used in compliance with the
described guidelines. Varying areas of application/
planned use must be approved by SAUTER in writing.

Guarantee

The guarantee is not valid following
• non-observation of our guidelines in the operating
instructions
• alteration to or opening of the device
• mechanical damage and damage caused by media,
liquids
• natural wear and tear
• inappropriate erection or electric installation
• overloading of the measurement equipment





Monitoring the test substances

The metrology features of the instrument and any possible
available adjusting weight must be checked at regular

Advertising
This manual is related to the following products: