5 calibration, Calibration – Maxim Integrated 78M6631 User Manual

Page 15

Advertising
background image

UG_6631_078

78M6631 Firmware Description Document

Rev 2

15

2.3.5 Calibration

As with any measurement system, there are also multiple sets of coefficients that are used to

compensate for system inaccuracies. Input registers for all coefficients can be manually modified and

saved to flash. Alternatively, high level calibration routines can be invoked. These routines automatically

determine the coefficients for common parameters and save them to flash memory. The different types of

compensation parameters include:

Voltage: gain and offset for voltage sense circuit

Current: gain and offset for current sense circuit

Phase: voltage-to-current phase offset introduced by transformer or filters in sense circuits

Temperature: offset for junction temperature at room temp

Temperature: coefficients for temperature curve

RMS and Power offset: fixed offsets for voltage and current and power at light loads

The calibration routines compensate for sensors and system inaccuracy. The new coefficients computed

during calibration can be stored in the on-chip flash.

In order to perform a calibration, an external stable AC source (Voltage/Current) and AC load are

required. The calibration routines have a target voltage and current (Register Itarget) used for all phases

to match. The target calibration voltage and current values are specified in the registers Vtarget and

Itarget.

A

Neutral

B

C

VC

VB

VA

IAP

IAN

IBP

IBN

ICP

ICN

V3

P3

A

A

Neutral

B

C

78M6631

AC SOURCE

AC

LOAD

AC

LOAD

AC

LOAD

NOTE: ONE OR MORE

PHASES CAN BE

CALIBRATED

SIMULTANEOUSLY

D

/Y

Figure 8. 3-Phase System Calibration Test Setup Example

Advertising