DATA LOGGER FUNCTION

The data logger provides ways to externally indicate or log all the incubator front panel displays at intervals
ranging from 1 to 99 minutes. The logger contains a real time clock/calendar to date and time stamp each
logged reading.

The logger indicates and logs all front panel display readings in single lines of text. This data is logged when
the present logging interval has expired or one of the monitored displays has changed its alarm state. Logged
data has a time and date field that corresponds with the logger

’s real time clock.

This logged data can be externally accessed through ports provided on the front of the incubator for serial and
parallel functions, as well as a BNC contact closure plug for extended alarm condition(s). The parallel port is
compatible with any Epson or IBM style printer. The serial port operates at 9600 baud with 8 data bits, 1 (one)
stop bit and no parity and connects with a computer through the USB.

6.1

Program the Data Logger: Select the parameter to adjust by pressing and holding the "MODE"
button pad until the left two display digits show "LP" indicating the logger has entered the set mode.
The parameter that can be changed will blink and its value can be changed by pushing the UP or
DOWN arrow pads.

6.2

Pressing and holding the MODE pad for approximately one second will cause the logger to step to the
next parameter. The table on the following page shows the indications for each parameter. If no
buttons are pressed for fifteen (15) seconds, the panel display will revert to the normal operation mode.

THE CONTROL PANEL DISPLAYS

SIX PARAMETERS, SHOWN BELOW, WHICH CAN BE CHANGED USING THE

CONTROL PANEL BUTTONS

FRONT PANEL MODE

LEFT TWO DIGITS

RIGHT TWO DIGITS

COLON

NORMAL OPERATION

SHOW HOURS OF CURRENT 24 HR

TIME

SHOW MINUTES OF CURRENT

TIME

BLINK

SET LOGGING PERIOD

SHOW "LP"

BLINK LOGGING INTERVAL IN

MINUTES. CAN BE VERIFIED

FROM 0-99 MINUTES .

OFF

SET HOURS

BLINKS HOURS PORTION OF

CURRENT TIME. CAN BE VARIED

FROM 00 TO 23

ARE BLANK

ON

SET MINUTES

ARE BLANK

BLINK MINUTES PORTION OF

CURRENT TIME. CAN BE VARIED

FROM 0 TO 59

ON

SET MONTH

BLINK CURRENT MONTH. CAN BE

VARIED FORM 1 TO 12

ARE BLANK

OFF

SET DATE

ARE BLANK

BLINK DAY OF MONTH. CAN BE

VARIED FROM 1 TO 31

OFF

CURRENT YEAR

"19" OR "20"

BLINK LOWER 2 DIGITS OF YEAR

OFF

6.3

The data logger clock has a back up battery. The recommended replacement is Panasonic 3 Volt, 165
mA # BR2325 type 1HB, 1HC, OR 1HG.