Shellab SCO10W-2 User Manual

Page 20

Advertising
background image

20

and see if heating light or safety light comes on
3/ check connections to sensor
4/ check calibration

– using independent certified reference

thermometer, follow instructions in sections 5.4 and 5.8

Unit will not heat up at all

1/ verify that controller is asking for heat by looking for controller light

if pilot light is not on continuously during initial start up there is a problem
with the controller
2/ check amperage

– amperage should be virtually at maximum rated

(data plate) amperage
3/ do all controller functions work?
4/ is the OTP set high enough?

– for diagnostics, should be fully

clockwise with the pilot light never on
5/ has the fuse/circuit breaker blown?

Indicated chamber temperature unstable

1/ ±0.1 may be normal
2/ is fan working?

– remove top panel and verify movement of cooling

fan in center of shadow box
3/ is ambient room temperature radically changing

– either door opening

or room airflow from heaters or air conditioning?

– stabilize ambient

conditions
4/ sensor miss-located, damaged, or wires may be damaged - check
mounts for control and OTP sensors, then trace wires between sensors
and controls
5/ calibration sensitivity

– call Customer Service

6/ OTP set too low

– be sure that OTP set point is more than 5 degrees

over desired Main set point; check if OTP pilot is on continuously; turn
controller knob completely clockwise to see if problem solved then follow
instructions in section 5.8.E for correct setting.
7/ electrical noise

– remove nearby sources of RFI including motors,

arcing relays or radio transmitters
8/ bad connection on temperature sensor or faulty sensor

– check

connectors for continuity and mechanical soundness while watching
display for erratic behavior; check sensor and wiring for mechanical
damage
9/ bad connections or faulty capacitor

– check connectors for

mechanical soundness and look for corrosion around terminals or signs
of arcing or other visible deterioration

Will not maintain set point

1/ assure that set point is at least 5 degrees over ambient room
temperature
2/ see if ambient is fluctuating

Display and reference thermometer don’t
match

1/ calibration error

– see section 5.4

2/ temperature sensor failure

– evaluate if pilot light is operating

correctly
3/ controller failure

– evaluate if pilot light is operating correctly

4/ allow at least two hours to stabilize
5/ verify that reference thermometer is certified

Can't adjust set points or calibration

1/ turn entire unit off and on to reset
2/ if repeatedly happens, call Customer Service

Calibrated at one temperature, but not at
another

This can be a normal condition when operating temperature varies
widely. For maximum accuracy, calibration should be done at or as
close to the set point temperature.

Advertising
This manual is related to the following products: