Co2 level – Shellab SCO10W-2 User Manual

Page 21

Advertising
background image

21

CO2 LEVEL

Overshoots set point but stabilizes -
display and Fyrite match

1/ turn set point up and down to see if solenoid valve works by feeling
and listening to valve
2/ recalibrate with Fyrite, see section 5.11 and section 7.0
3/ fan not operating correctly
a- fan motor stopped
b- fan blade fell off
c- wrong fan blade installed or mounted backwards
4/ wrong restrictor installed or missing altogether
5/ tank pressure too high, see section 5.9.A
6/ CO2 sensor partially plugged with dirt or condensation
7/ regulator set wrong, see section 5.9
8/ incubator too heavily loaded
9/ incubator being operated without shelving
10/ CO2 attached to sample port instead of “CO2 IN” fitting
12/ if display and Fyrite does not match, see Troubleshooting Co2.

Overshoots set point and continues to rise
- display and Fyrite match

1/ debris in solenoid causing it to leak continuously
2/ solenoid failed while open
3/ unit miss-wired
4/ controller output failed or shorted
5/ CO2 sensor or interface failure
6/ CO2 sensor plugged by debris or condensation

Rises very slowly

1/ restrictor partially plugged
2/ filter overly dirty or partially plugged
3/ CO2 tank regulator set too low, see section 5.9
4/ hose kinked or leaking
5/ poor door seal
6/ CO2 tank contains mixed gas, not 100% medical grade CO2

Never rises

1/ CO2 tank empty
2/ solenoid failed while closed
3/ CO2 controller output failed while open
4/ restrictor plugged
5/ CO2 hose blockage
6/ CO2 filter plugged
7/ set point is at 0.0 and has not been reset, see section 5.3 and 5.4

Display and Fyrite reading do not match