Calibrate the temperature display, Operation (continued) – Shellab SRI20PF User Manual

Page 21

Advertising
background image

21 |

P a g e

OPERATION (CONTINUED)

C

ALIBRATE THE

T

EMPERATURE

D

ISPLAY

Allow the incubator at least 24 hours to stabilize at its operational set point before performing a
calibration. Always use a certified temperature sensing reference device that is regularly calibrated
to 0.1°C by a third party to conduct temperature verifications or calibrations. For best results use a
remote sensing device.

Part 1: Calibrate the Day Time Temperature

1. Place the incubator Manual Mode, with the Day Time Temperature indicator LED

illuminated.

2. Place the temperature sensor of a calibrated reference device inside the incubator as close

as possible to the

chamber’s geometric center.

Note:

A thermocouple sensor probe’s sleeve may be taped to the shelving, as long as the exposed
copper end is 2 inches (5cm) above the shelf. An exposed sensor probe in direct contact
with the shelving may experience heat sinking, which can result in an inaccurate temperature
reading.

3. Allow the temperature to re-stabilize after closing the door. The reference device

’s

temperature measurement should not change for at least one (1) hour after a significant
disruption for the chamber to be considered stabilized.

4.

If the reference device’s and the incubator’s temperature readings are the same, or the
difference between the two falls within the acceptable range of your laboratory protocol
after one (1) hour of stability, the incubator is calibrated.

5. If there is a difference between the two (2) readings, and that offset falls outside your

laboratory protocol

’s acceptable range, adjust the incubator’s temperature display to match

the

reference device’s reading. See the next step. If the door was briefly opened to take a

reading, wait fifteen minutes for the temperature to stabilize before correcting for the offset.

6. Press and hold both the UP and DOWN arrows simultaneously. The Temperature Display

will begin flashing and display the current temperature value. Adjust the current value
until it matches the reference device’s temperature reading. This will enter the offset into
the unit’s temperature controller board.

7.

After matching the unit’s current value to that of the reference device, wait five seconds.
The incubator’s Temperature Display will cease flashing, and the incubator will now begin
heating or cooling to compensate for the entered temperature offset.

8. Allow the incubator one (1) hour to stabilize after reaching the new temperature.

9.

Compare the reference device’s reading with the incubator’s. If the reference device’s and
the incubator’s temperature readings are the same or fall within the range of your
laboratory protocol, the incubator is now calibrated.

10. If the two readings are not the same or fall outside your laboratory protocol, repeat steps 5

– 8.

11. If the temperature readings of the incubator and the reference device still fall outside your

laboratory protocol after three calibration attempts,

contact Sheldon Technical Support

for assistance.

Advertising
This manual is related to the following products: